Wooden crate: Open the shipping crate
-
Wooden crate: Remove the wooden panel to get access to the floor panel
-
NEW Cardboard crate: If your system is shipped in the new cardboard crate, simply remove the lid and carefully cut open the cardboard (the crate is single use).
-
-
-
-
Click the connector at the roof for each panel in the corresponding socket at the top of the rear, left and right side panel
-
On the rear panel, push the grey flat cable carefully through the opening in the roof, and connect it to the CPU module (there is only one open socket on the CPU module).
-
-
-
Install the door (be careful: do NOT lift the door by the plastic door handle!)
-
If the metal pin is not in the door yet, you have to install it at the bottom of the door before proceeding.
-
Align the metal pin at the bottom of the door into the teflon washer at the left side of the floor (see red marker in the photo)
-