Major leakage erro please help
Hello guys
This error message appears and no tidal volume
Please help
Is dit een goede vraag?
Hello guys
This error message appears and no tidal volume
Please help
Is dit een goede vraag?
@zorro979 you most likely have issues witht he Tidal volume based on the leakage. I suggest you download the service manual from here and perform the Valve test which goes through the leakage checks. There are to many possibilities which require multiple tests and checkpoints.
Was dit antwoord nuttig?
Check for leakage of the endotracheal tube or LMA whatever you are using. If not the normal size, it will alarm leakage. What's the level of your oxygen? If more than 3L/Min , it will alarm leakage. Please set your PEEP to either 4 or 5.. do your machine check and correct the Fio2 21% and 100%.
Was dit antwoord nuttig?