  1. Hamilton-C3 Oxygen Cell Replacement, Oxygen Cell: stap 1, afbeelding 1 van 1
    • At the base of the ventilator housing (see diagram) is a removable cover at the far left of the panel. Open the cover by depressing the release button to expose the existing oxygen cell in its socket (see 1).

  2. Hamilton-C3 Oxygen Cell Replacement: stap 2, afbeelding 1 van 1
    • Pull gently to remove the electrical lead (see 2).

    • When reassembling, apply the electrical lead to the terminal until the lead locks or latches onto the terminal. Ensure that a solid contact is made with the electrical lead and take care to prevent the lead from becoming dislodged or disconnected during the installation.

  3. Hamilton-C3 Oxygen Cell Replacement: stap 3, afbeelding 1 van 1
    • Rotate the oxygen cell counterclockwise to remove the cell from its socket (see 3).

  4. Hamilton-C3 Oxygen Cell Replacement: stap 4, afbeelding 1 van 1
    • To reassemble your device, follow the instructions in reverse order.