Ga door naar hoofdinhoud

Newport HT70 Pump Assembly Replacement

Medische apparatuur
Noch iFixit, noch deze reparatiehandleiding worden onderschreven door of zijn gelieerd aan de fabrikant van de desbetreffende medische apparatuur.

Wat je nodig hebt

  1. Newport HT70 Pump Assembly Replacement, Pump Assembly: stap 1, afbeelding 1 van 1
    • Position the base assembly so that you may access it.

    • Using the #1 Phillips screwdriver, unscrew the four screws (SCR3219P in the photo) and the two screws (SCR3232P) securing the base plate.

  2. Newport HT70 Pump Assembly Replacement: stap 2, afbeelding 1 van 1
    • The base plate will then come apart from the rest of the device with the pump assembly attached to it, as depicted in the following photo.

    • Inspect pump assembly for any damage or looseness.

  3. Newport HT70 Pump Assembly Replacement: stap 3, afbeelding 1 van 1
    • If damage is found, replace the pump assembly by removing each tube individually and replacing as needed.

    • You may not need to replace the entire assembly if parts of it are still intact.

Conclusie

To reassemble your device, follow these instructions in reverse order.