SPECIFICATIONS
Selectable Loads and Ranges (Monopolar and Bipolar)
| Load (Ohms) | Range (Watts) |
|---|---|
| 50 | 50 |
| 100 | 100 |
| 100 | 300 |
| 200 | 300 |
| 300 | 300 |
| 400 | 500 |
| 500 | 500 |
Table Caption: This table lists various electrical loads in Ohms and their corresponding power ranges in Watts. It shows that for some load values, multiple power ranges are possible.
Isolation Test Range:
Current Range:
0 -50 Watts with 200-Ohm Load from Patient Plate, Active Electrode, and Chassis
0 -500 mA
Return Fault:
150 picofarads
Frequency Response : 0-5MHz
Frequency Response: (Filter Pushed) 0 - 50 kHz
Load Accuracy: + 2% at 50 ohms + 1.5% at 100 ohms + 1% 200 to 500 ohms
Meter Accuracy:
Power Output Accuracy:
-
2%
-
5% (Calibrated in Watts)
REM Test: Normal 24.6 Ohm Resistance, 137.0 Ohms with REM Switch Activated
Oscilloscope:
Power Requirements:
Measuring Technique:
Weight:
Isolated (Uncalibrated)
None
RF Thermocouples
5.4
kg
(12
Ibs)
Dimensions: 28 cm X 23 cm X 18 cm -(11"X 9.1" X7")
"