Ventilator ABV - A - U

Service Manual

Contens

1. DESCRIPTION OF CONSTRUCTION AND FUNCTION4
1.1 Complete scheme4
1.2 Respirator with patient unit5
Only ABV-U5
1.3 Adjustment of parameters at the respirator6
1.3.1 Change of inspiration flow at the flow-control valve6
1.3.2 Frequency adjustment6
1.3.3 Breathing/time ratio (inspiration/expiration)6
1.3.4 Pressure monitor6
1.3.5 Control limit-value IPPB7
1.4 Patient component ABV-A8
1.4.1 PEEP-adjustment8
1.4.2 PLATEAU adjustment8
1.4.3 Change lever Respirator / manual8
1.4.4 Ejektor / GAS EVAC8
1.5 Patient unit for paediatric respiration8
2. FUNCTION SCHEME9
2.1 Technical course of events during inspiration9
2.2 Technical course of events during expiration10
3. SERVICING OF BASIC APPARATUS11
3.1 Electronic11
3.2 Pneumatic12
4. SIGHT CHECK12
5. PERFORMANCE CHECK12
5.1 Self-test of respirator12
5.2 Errorcodes / Operation - Display13
5.3. Operating modeCMV13
5.4 Operating mode IPPV14
5.5 Pressure alarms14
6. CHECK OF ELECTRICAL SAFETY (STK)15
7. PATIENT UNIT ABV-A16

Table Caption: This table of contents outlines a technical manual for a respirator, detailing its construction, function, servicing, performance checks, and safety procedures. It covers both the basic apparatus and patient-specific components, including those for pediatric use.

6

6

6

6

7

8

8

8

8

8. CLEANING & STERILISATION17
8.1 Ventilator ABV-A17
8.2 Patient unit17
9. ERROR-FLOW DIAGRAM19
9.1 Respirator19
9.2 Pressure monitor19
10. SPECIFICATION SHEET20
10.1 Ventilator module ABV - A20
10.2 Patient component21
11. SPARE - PART - LIST22
12. MAINTENANCE23

Table Caption: This table appears to be a table of contents or index, listing sections of a document related to cleaning, error diagrams, specifications, spare parts, and maintenance, along with their corresponding page numbers.

1. Description of construction and function

1.1 Complete scheme

The respirator module is equipped with a pneumatic and with an electronic control unit (respirator ABV-A and patient unit), and with an electronic control monitor for breathing pressure.

The electronic of the respirator module is supplied with line voltage (230V/50 c.p.s.), while the pneumatic unit requires a propellant gas in form of compressed air (3 - 5 bar). The respirator ABV-A is designed for the anaesthesia respiration of children and adults in a semiclosed system.

The patient unit can be exchanged quickly due to its quick-release fastener and meets the highest requirements for hygiene because of its easy dismountability (partly autoclavable with a temperature of up to 134 ° C).

General description

The respirator ABV-A has to be supplied with compressed air as propellant gas with a pressure of 3 - 5 bar (1). The various respiration parameters can be adjusted at the respirator:

1.2 Respirator with patient unit

The respirator ABV-A is designed for controlled anaesthesia respiration with state-of-the-art technology.

The respirator works in a pressure-gas