Ga door naar hoofdinhoud

The Ohmeda Excel 210 is an anesthesia machine with powerful ventilation and monitoring capabilities. It can accommodate three gases and two vaporizers.

1 Vragen Bekijk alle

I have Ohmeda 210 Ventilator , but it’s ventilator is not working

I have Ohmeda 210 Ventilator , but it’s ventilator is not working I have Ohmeda 210 Ventilator , but it’s ventilator is not working because of its supply issue and I think its supply is changed and I don’t know it’s voltage.. please help

Beantwoord deze vraag Dit probleem heb ik ook

Is dit een goede vraag?

Score 1

1 Opmerking:

[code]

[video|4948][image|3031385][image|3031387][image|3031388][image|3031384][image|3031383]

[/code]

door

Voeg een opmerking toe

1 Antwoord

Here are some tips for troubleshooting an Ohmeda 210 ventilator that is not working due to a potential power supply issue:

  • The Ohmeda 210 can operate on either AC or DC power. For AC, it requires a power input of 100-240VAC at 50/60Hz. For DC, it requires 11-15VDC.
  • First, verify the ventilator is being powered - check that it is plugged in, the outlet has power, and any circuit breakers are not tripped.
  • Inspect the power cable for damage and ensure it is securely plugged into the ventilator. Replace if damaged.
  • Open the ventilator case and check that the internal power connector is properly attached to the main circuit board. Reseat if needed.
  • Using a multimeter, check the voltage at the power input terminals or connectors on the main board. It should match the nameplate voltage for AC or DC operation.
  • If the input voltage is correct, the issue may be a failed internal power supply module. This will need replacement.
  • Also, inspect for any burnt/damaged components on the main board that could indicate a power surge issue.
  • Check error codes for any power or supply-related alerts.
  • As a temporary solution, try using an external DC power supply matched to the unit's input specs.

Document any error conditions and double-check the input voltage/connections to determine if the issue is internal or external to the ventilator.

Was dit antwoord nuttig?

Score 0

1 Opmerking:

Please check its pictures, I think its power supply is changed so we can’t replace it when we won’t exactly know the voltage properly

door

Voeg een opmerking toe

Voeg je antwoord toe

Alpha Professionals zal eeuwig dankbaar zijn.
Weergavestatistieken:

Afgelopen 24 uren: 2

Afgelopen 7 dagen: 3

Afgelopen 30 dagen: 8

Altijd: 33