Ga door naar hoofdinhoud

Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement

Wat je nodig hebt

  1. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: stap 1, afbeelding 1 van 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: stap 1, afbeelding 2 van 2
    • Unplug the AC power connection cable.

  2. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 2, afbeelding 1 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 2, afbeelding 2 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 2, afbeelding 3 van 3
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

  3. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 3, afbeelding 1 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 3, afbeelding 2 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 3, afbeelding 3 van 3
    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

  4. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 4, afbeelding 1 van 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 4, afbeelding 2 van 2
    • Grasp the temperature probe and pull it up to remove it from the monitor.

  5. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 5, afbeelding 1 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 5, afbeelding 2 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 5, afbeelding 3 van 3
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

  6. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 6, afbeelding 1 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 6, afbeelding 2 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 6, afbeelding 3 van 3
    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

  7. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 7, afbeelding 1 van 1
    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

  8. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 8, afbeelding 1 van 1
    • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

  9. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 9, afbeelding 1 van 2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 9, afbeelding 2 van 2
    • Detach any accessory USB cables from USB ports on the monitor.

  10. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 10, afbeelding 1 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 10, afbeelding 2 van 3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: stap 10, afbeelding 3 van 3
    • Slide the monitor off of the stand tray.

Conclusie

To reassemble your device, follow these instructions in reverse order.

Sam Goldheart

Lid sinds: 10/18/12

465.762 Reputatie

538 handleidingen geschreven

Team

iFixit Lid van iFixit

Staff

135 Leden

16.817 handleidingen geschreven

0 opmerkingen

Voeg opmerking toe

Weergavestatistieken:

Afgelopen 24 uren: 0

Afgelopen 7 dagen: 2

Afgelopen 30 dagen: 4

Altijd: 863