Weinmann VENTImotion and VENTIlogic Display Board Replacement

Medische apparatuur
Noch iFixit, noch deze reparatiehandleiding worden onderschreven door of zijn gelieerd aan de fabrikant van de desbetreffende medische apparatuur.

Wat je nodig hebt

  1. Weinmann VENTImotion and VENTIlogic Display Board Replacement, Display Board: stap 1, afbeelding 1 van 2 Weinmann VENTImotion and VENTIlogic Display Board Replacement, Display Board: stap 1, afbeelding 2 van 2
    • Open the device by following these steps:

    • Place the device on a non-slip surface with the top facing downwards

    • Remove the filter cassette from the top part of the housing.

    • Undo and remove the 6 screws (34).

    • Open the bottom of the housing (35) away to the side.

    • Undo the connecting cables for the humidifier (42) and the alarm (43).

    • Make sure the power supply plug is disconnected.

  2. Weinmann VENTImotion and VENTIlogic Display Board Replacement: stap 3, afbeelding 1 van 1
    • Disconnect the following:

    • encoder connector (82).

    • connectors of interfaces (84) and (85) off the control board.

  3. Weinmann VENTImotion and VENTIlogic Display Board Replacement: stap 4, afbeelding 1 van 1
    • Disconnect the ribbon cable of the display and the fascia film.

    • Pull the latch out until you feel resistance. You can then pull out the ribbon cable.

    • If the latch has not been opened correctly, the ribbon cable may be damaged when it is pulled out.

  4. Weinmann VENTImotion and VENTIlogic Display Board Replacement: stap 5, afbeelding 1 van 1
    • Pull flow measurement hoses (56) off the flow sensor.

    • Pull pressure measurement hoses (57) off the pressure sensors.

  5. Weinmann VENTImotion and VENTIlogic Display Board Replacement: stap 6, afbeelding 1 van 1
    • Undo the four screws (51) and remove control board (50).

  6. Weinmann VENTImotion and VENTIlogic Display Board Replacement: stap 7, afbeelding 1 van 1
    • Remove spring (54) from the control board and remove battery (52) (see "7.9 Replace battery on control board" on page 42).

    • Remove display

  7. Weinmann VENTImotion and VENTIlogic Display Board Replacement: stap 9, afbeelding 1 van 1
    • Insert a new display (49).

    • The display may not tilt and must be straight in the guides.

Conclusie

To reassemble your device, follow these instructions in reverse order.

Cameron Young

Lid sinds: 04/22/19

174 Reputatie

1 Handleiding geschreven

Team

Cal Poly, Team S7-G19, Paton Spring 2020 Lid van Cal Poly, Team S7-G19, Paton Spring 2020

CPSU-PATON-S20S7G19

3 Leden

4 handleidingen geschreven

0 opmerkingen

Voeg opmerking toe

Weergavestatistieken:

Afgelopen 24 uren: 0

Afgelopen 7 dagen: 0

Afgelopen 30 dagen: 1

Altijd: 88