    • Check to see if the exhalation filter assembly is installed, if it is, remove the exhalation filter assembly.

    • Rotate the metal locking lever on the lower right of the ventilator body forward to an open position.

    • Remove the exhalation filter assembly from the ventilator body by pulling straight down.

    • Using a long Phillips screwdriver, remove the Phillips screw located at the top of the exhalation filter assembly well, if applicable.

    • Using a long Phillips screwdriver, remove the Phillips screw located at the exhalation port marked EXH, if applicable.

    • Using a long Phillips screwdriver, remove the Phillips screw from the outside of the plastic cover.

    • Using a long Phillips screwdriver, remove the Phillips screw on the font left wall.

    • Using a long Phillips screwdriver, remove the Phillips screw on the right side of the wall.

    • Set the plastic cover aside.

    • Using a long Phillips screwdriver, remove the 3 Phillips screws on the metal re-enforcement cover, if applicable, and set aside.

    • Remove the Exhalation Corner Assembly.

Conclusie

To reassemble your device, follow the steps in the reverse order.

Één andere persoon heeft deze handleiding voltooid.

Alden Glassey

Lid sinds: 04/19/18

498 Reputatie

3 handleidingen geschreven

Team

Cal Poly, Team S1-G1, Zimmerman Spring 2020 Lid van Cal Poly, Team S1-G1, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G1

5 Leden

6 handleidingen geschreven

0 opmerkingen

Voeg opmerking toe

Weergavestatistieken:

Afgelopen 24 uren: 0

Afgelopen 7 dagen: 5

Afgelopen 30 dagen: 6

Altijd: 98