This guide deals with an important prerequisite, the removal of the flow sensor. Once removed, many more components can be accessed on the top of the e500 ventilator. This step is very simple, but must be done carefully as there are a lot of exposed parts in the process. For more information refer to page 43 of the service manual.

  1. Remove the three screws and washers that secure the flow sensor cover to the pneumatic box assembly. Retain the hardware for re-assembly.
    • Remove the three screws and washers that secure the flow sensor cover to the pneumatic box assembly. Retain the hardware for re-assembly.

  2. Remove the flow sensor cover.
    • Remove the flow sensor cover.

Conclusie

To reassemble your device, follow these instructions in reverse order.