If you need to access important parts associated with the Newort e500, removal of the exhalation module is an integral step. This step serves as a common pre-requisite to completing other guides. Make sure you are careful with the outlet boot as it is fragile. For more information refer to page 42 of the service manual.

  1. Pull up and remove the flexible rubber exhalation outlet boot (push fit).
    • Pull up and remove the flexible rubber exhalation outlet boot (push fit).

  2. Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.
    • Disconnect the tubes from both sides of the flow sensor and the two tubes from the exhalation valve.

  3. Pull the exhalation module straight up to remove it from the GDU.
    • Pull the exhalation module straight up to remove it from the GDU.

Conclusie

To reassemble your device, follow these instructions in reverse order.