Ga door naar hoofdinhoud

Hamilton Medical Ventilator Replacing the LCD

Medische apparatuur
Noch iFixit, noch deze reparatiehandleiding worden onderschreven door of zijn gelieerd aan de fabrikant van de desbetreffende medische apparatuur.

Wat je nodig hebt

  1. Hamilton Medical Ventilator Replacing the LCD, LCD: stap 1, afbeelding 1 van 1
    • First open the main enclosure by removing the four hexagonal-drive (Allen) screws at the rear of the enclosure.

  2. Hamilton Medical Ventilator Replacing the LCD: stap 2, afbeelding 1 van 1
    • Unscrew the control board mounting plate.

  3. Hamilton Medical Ventilator Replacing the LCD: stap 3, afbeelding 1 van 1
    • “Hinge” back the mounting plate, complete with control board. This exposes the LCD.

  4. Hamilton Medical Ventilator Replacing the LCD: stap 4, afbeelding 1 van 1
    • Unplug the cable from the GMP adaptor board from the LCD.

  5. Hamilton Medical Ventilator Replacing the LCD: stap 5, afbeelding 1 van 1
    • Unplug from the dc/ac board, the cable to the interface board.

  6. Hamilton Medical Ventilator Replacing the LCD: stap 6, afbeelding 1 van 1
    • Unscrew the four hexagonal-drive (Allen) screws securing the two brackets on which the LCD is mounted.

  7. Hamilton Medical Ventilator Replacing the LCD: stap 7, afbeelding 1 van 1
    • Lift out the complete assembly.

  8. Hamilton Medical Ventilator Replacing the LCD: stap 8, afbeelding 1 van 1
    • Unplug both of the lights from the dc/ac board.

  9. Hamilton Medical Ventilator Replacing the LCD: stap 9, afbeelding 1 van 1
    • Remove the LCD from the two brackets, by removing the black hexagonal-drive (Allen) screws on the LCD. NOTE: Do not remove the silver hexagonal-drive (Allen) screws on the SIDE of the LCD.

  10. Hamilton Medical Ventilator Replacing the LCD: stap 10, afbeelding 1 van 1
    • When fitting a new display, there must be an offset of approximately 2 mm between the plastic sealing strips on the LCD, and the brackets.

  11. Hamilton Medical Ventilator Replacing the LCD: stap 11, afbeelding 1 van 1
    • If necessary, adjust this gap using the silver screws on the side of the LCD.

    • The surface of the LCD must be clean. Wipe it free of finger marks and dust with an antistatic paper towel.

  12. Hamilton Medical Ventilator Replacing the LCD: stap 13, afbeelding 1 van 1
    • The cable from the GMP to the LCD must be in the cable grip on the control board mounting plate.

    • When screwing the LCD assembly into GALILEO, you must align it with the glass screen in GALILEO. Check this by looking at the front of GALILEO.

    • Turn on GALILEO, and make sure that the LCD lights up, and is properly aligned.

Conclusie

To reassemble your device, follow these instructions in reverse order.

Sofya Shatalova

Lid sinds: 04/07/20

229 Reputatie

1 Handleiding geschreven

Team

Cal Poly, Team S7-G26, Paton Spring 2020 Lid van Cal Poly, Team S7-G26, Paton Spring 2020

CPSU-PATON-S20S7G26

3 Leden

4 handleidingen geschreven

0 opmerkingen

Voeg opmerking toe

Weergavestatistieken:

Afgelopen 24 uren: 0

Afgelopen 7 dagen: 0

Afgelopen 30 dagen: 2

Altijd: 244