ZT-703S frozen UI/buttons after oscilloscope mode + “RAM” USB issue —
Hi everyone,
I’m looking for help with a possibly soft-bricked ZOYI/Zotek ZT-703S oscilloscope multimeter.
Symptoms:
- The device powers on normally
- It immediately enters oscilloscope mode
- The oscilloscope itself still works and reads signals correctly
- However, all front buttons become unresponsive in the main UI
What is interesting:
- The buttons DO work in bootloader/USB mode
- POWER + F1 during startup successfully enters USB mode
- F2 successfully exits USB mode
I also have the known “RAM” USB issue:
- The device appears as “RAM” over USB
- Reduced storage size is shown
- Larger firmware files no longer fit
I already tried multiple official firmware versions, but the issue persists.
After researching EEVBlog and the abandoned open-zt703 project, I suspect:
- corrupted persistent settings/filesystem,
- corrupted internal flash state,
- or a firmware/UI crash when restoring oscilloscope settings at boot.
Important findings:
- The STM32 SWD interface appears to still be enabled
- The open-source project documentation mentions:
- full chip erase,
- filesystem recreation,
- settings reset,
- and recovery through ST-Link + STM32CubeProgrammer
At this point I’m considering buying an ST-Link V2 and attempting recovery through SWD.
Before I do that, I wanted to ask:
- Has anyone successfully recovered a ZT-703S from this exact state?
- Is there a known way to reset only the settings/filesystem area?
- Does a full chip erase also destroy the original bootloader permanently?
- Has anyone dumped the original flash/bootloader?
Any advice would be greatly appreciated.
Video demostration https://www.youtube.com/shorts/B2H7HaynO...